FDA Adverse Event Injury Summary report: N

NFLEX STRAIG 1SEGM L150 2U PCU/TAV-ELI

MDR report key: 2955416 · Received February 11, 2013

Report

Report Number
2520274-2013-00830
Event Type
Injury
Date Received
February 11, 2013
Report Date
January 14, 2013
Manufacturer
SYNTHES USA
Product Code
NQP
PMA / PMN Number
K061774
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS THE SUBJECT DEVICE WAS EVALUATED AT (B)(4). THE CHEMICAL COMPOSITION OF BOTH RODS WAS TESTED BY (B)(4). THE RODS WERE FOUND TO BE MADE OF TITANIUM ALLOY ((B)(4)). A SMALL PORTION OF EACH ROD WAS SECTIONED AND PREPARED FOR METALLOGRAPHIC EXAMINATION INTO A TRANSVERSE AND LONGITUDINAL MICRO-SECTION. THE TRANSVERSE MICRO-SECTIONS OF BOTH RODS SHOWED A HOMOGENEOUSLY SHAPED ALPHA AND BETA MICROSTRUCTURE OF A1 (ISO 20160). ALSO, THE LONGITUDINAL MICRO-SECTIONS SHOWED A TYPICAL MICROSTRUCTURE FOR THIS TYPE OF MATERIAL AND ARE FREE OF ALPHA-CASE. THE METALLOGRAPHIC EXAMINATION SHOWED THAT THE MICROSTRUCTURE OF THE RODS IS COMPLYING WITH THE INTERNATIONAL STANDARDS ISO 5832-3 AND ASTM F136. HARDNESS MEASUREMENTS WERE CARRIED OUT USING A VICKERS INDENTER. THE AVERAGE HARDNESS OF ROD 1 WAS FOUND TO BE (B)(4) WHICH EQUATES TO AN APPROXIMATE TENSILE STRENGTH OF (B)(4), THE AVERAGE HARDNESS OF ROD 2 WAS FOUND TO BE (B)(4) WHICH EQUATES TO AN APPROXIMATE TENSILE STRENGTH OF (B)(4). THE REQUIREMENT OF THE TENSILE STRENGTH ACCORDING TO THE INTERNATIONAL STANDARD ISO 5832-3 IS A MINIMUM VALUE OF 860 MPA. THEREFORE, THE MATERIAL OF THE RODS MEETS AND EXCEEDS THE REQUIREMENTS OF THE INTERNATIONAL STANDARDS ISO 5832-3 AND ASTM F136. THE DIMENSIONS OF THE RODS (AS FAR AS MEASURABLE) WERE CHECKED USING A DIGITAL SLIDING CALIPER. THE RESULTS OF THE OUTER DIAMETER WERE FOUND TO BE IN COMPLIANCE WITH THE TECHNICAL DRAWING AND AO/ASIF SPECIFICATIONS. THE RECEIVED NFLEX RODS ARE MADE OF TWO DIFFERENT DESIGN VERSIONS. BECAUSE OF THE INTRA-OPERATIVE SHORTENED ENDS OF THE RODS IT IS NOT POSSIBLE TO TRACE THE ARTICLE NUMBER RESPECTIVELY TO ASSIGN THE RODS TO A DESIGN VERSION. AFTER BREAKING, THE FRAGMENTS RUBBED AGAINST EACH OTHER CAUSING PARTIALLY SLIGHT DESTRUCTION OF THE FRACTURE SURFACE (SHINY SURFACE, ABRADED AREAS). WEAKLY VISIBLE LINES OF REST WERE OBSERVED AND ARE AN INDICATION OF A FATIGUE FRACTURE. WHEN EXAMINING THE CAUDAL FRACTURE SURFACES USING THE SCANNING ELECTRON MICROSCOPE (SEM), THE INITIAL FRACTURE AREAS AND THE FRACTURE BEHAVIOR WERE IDENTIFIED. ON BOTH RODS THE CRACK INITIATED ON THE SURFACE OF THE ROD AND RAN THROUGH THE MATERIAL. A PATTERN OF RIPPLES OR FATIGUE STRIATIONS WAS OBSERVED AT THE CRACK PROPAGATION ZONES. EACH STRIATION REPRESENTS THE SUCCESSIVE POSITION OF AN ADVANCING CRACK FRONT. THE PRESENCE OF THESE STRIATIONS IS A CLEAR INDICATION OF A FATIGUE PROCESS. THE FRACTURE SURFACE OF ROD 1 SHOWED A PRIMARY AND A SECONDARY CRACK INITIATION ZONE INDICATING TWO-SIDED BENDING LOADS. A FATIGUE CRACK AT THE SECONDARY CRACK INITIATION ZONE IS DOCUMENTED. THE PORTION OF FORCED FRACTURE WITH DIMPLES OF APPROXIMATELY 75% IS VERY HIGH AND INDICATES HIGH DYNAMIC LOADS. THE FRACTURE SURFACE OF ROD B ALSO SHOWED FATIGUE CRACKS AT THE CRACK INITIATION ZONE. THE PORTION OF FORCED FRACTURE WITH DIMPLES WITH APPROXIMATELY 85% IS ALSO VERY HIGH. SEM OBSERVATIONS AND FINDINGS INDICATE THAT THE ROD FAILURES WERE CAUSED BY FATIGUE AND OVERLOAD. THE FAILURE OF THE INVESTIGATED NFLEX RODS WAS DUE TO DYNAMIC BENDING WHICH LED TO THE MATERIAL FATIGUE. BASED ON THE TOPOGRAPHY OF THE FRACTURE SURFACES WE CAN CONCLUDE THAT THE INVESTIGATED IMPLANT HAD TO ABSORB AND NEUTRALIZE FORCES THAT HAVE BEEN MUCH GREATER THAN THE TOLERABLE LOAD. THE INCREASE OF WEIGHT OF THE PATIENT ((B)(6)) HAS PLAYED A CERTAIN ROLE TOO. A FAILURE RESULTING FROM EITHER A MATERIAL DEFECT OR THE MANUFACTURING PROCESS CAN BE EXCLUDED.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: (B)(6) 2011 THE PATIENT WAS IMPLANTED WITH TWO NFLEX RODS. POST-OPERATIVELY THE RODS BROKE AND ON (B)(6) 2012 WERE REMOVED. THE ETIOLOGY FOR THE BREAK WAS UNCLEAR. HOWEVER, IT WAS NOTED THE PATIENT RECENTLY HAD A SIGNIFICANT INCREASE IN BODY MASS FROM A BMI OF APPROXIMATELY (B)(6). IT WAS REPORTED BOTH RODS HAD COMPLETELY FRACTURED AT THE DYNAMIC RIGID ROD INTERFACE. THE PATIENT WAS REVISED WITH STANDARD CLICK X RODS. THE LOT NUMBER FOR THE IMPLANTED NFLEX RODS WAS NOT VISIBLE BECAUSE THE RODS WERE CUT FOR LENGTH ADAPTION.

Description of Event or Problem · 1

THIS REPORT IS 1 OF 2 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58841 NFLEX STRAIG 1SEGM L150 2U PCU/TAV-ELI ROD NQP SYNTHES USA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention