FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2955374 · Received February 11, 2013

Report

Report Number
9614453-2013-00289
Event Type
Injury
Date Received
February 11, 2013
Date of Event
December 7, 2012
Report Date
December 14, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL PRODUCT WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY SHOWED THE BATTERY INDICATOR SIGNIFYING THAT IT IS TIME FOR DEVICE REPLACEMENT. (B)(4).

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE U.S., HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. (B)(4).

Description of Event or Problem · 1

IT WAS ALSO REPORTED THAT THE PATIENT HAD SYMPTOMS OF DYSPNEA.

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE REACHED RECOMMENDED REPLACEMENT TIME AFTER TWO YEARS DUE TO HIGH OUTPUTS ON THE ATRIAL CHANNEL. PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND WAS REPLACED. THE PATIENT IS A PARTICIPANT IN THE OPTILINK STUDY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59770 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D234TRK

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Hospitalization| R