FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2955365 · Received February 11, 2013

Report

Report Number
2649622-2013-01430
Event Type
Injury
Date Received
February 11, 2013
Report Date
December 6, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY REMOTE TRANSMISSION THE DAY AFTER IMPLANT, INTERMITTENT VENTRICULAR SENSING WITHOUT MARKERS. IT WAS FURTHER NOTED THE VENTRICULAR LEAD HAD HIGH THRESHOLDS. REPROGRAMMING WAS PERFORMED AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59513 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00092 YR Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR (IPG)