REVO MRI SURESCAN
Report
- Report Number
- 3004209178-2013-02064
- Event Type
- Injury
- Date Received
- February 11, 2013
- Date of Event
- December 15, 2012
- Report Date
- March 20, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWP
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT WITHIN A COUPLE DAYS OF IMPLANT, THE PATIENT WAS EXPERIENCING PROLONGED PAUSES, AND THE RIGHT VENTRICULAR (RV)LEAD EXHIBITED A LOSS OF CAPTURE AND WAS FOUND TO HAVE DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, THE PHYSICIAN WAS UNABLE TO TIGHTEN THE RV SETSCREW ON THE EXISTING DEVICE. AS A RESULT, THE DEVICE WAS EXPLANTED AND REPLACED AS WELL. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT WITHIN A COUPLE DAYS OF IMPLANT, THE PATIENT WAS EXPERIENCING PROLONGED PAUSES, AND THE RIGHT VENTRICULAR (RV)LEAD EXHIBITED A LOSS OF CAPTURE AND WAS FOUND TO HAVE DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, THE PHYSICIAN WAS UNABLE TO TIGHTEN THE RV SETSCREW ON THE EXISTING DEVICE. AS A RESULT, THE DEVICE WAS EXPLANTED AND REPLACED AS WELL. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59078 | REVO MRI SURESCAN | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | RVDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR | Hospitalization| L| R | 5086MRI X2 IMPLANTABLE PACING LEADS |