FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2955333
·
Received February 11, 2013
Report
- Report Number
- 2649622-2013-01368
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- December 12, 2012
- Report Date
- March 5, 2013
- Manufacturer
- MPRI
- Product Code
- NVY
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY #(B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE PRIMARY ANALYSIS FINDING NOTED NO ANOMALIES WERE FOUND.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD HAD OVERSENSING AND HIGH IMPEDANCE. THE HELIX DID NOT EXTEND OR RETRACT EASILY. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59446 | SPRINT QUATTRO SECURE | DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES | NVY | MPRI | 6947M62 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |