FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2955328 · Received February 11, 2013

Report

Report Number
2649622-2013-01484
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
December 27, 2012
Report Date
December 27, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS 4076 IMPLANTABLE PACING LEAD - (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT T-WAVE OVERSENSING (TWOS). IT WAS DETERMINED THAT THE DEVICE ALGORITHM WOULD ACTIVATE WHEN THE CORRECT NUMBER OF INTERVALS TO DETECT (NID) WOULD BE MET, AND AS SUCH, NO PROGRAMMING CHANGES WERE MADE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58974 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M55

Patients

Seq Age Sex Outcome Treatment
1 00035 YR D314DRM IMPLANTABLE DEFIBRILLATOR