ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-01682
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Report Date
- January 26, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: AN "EZPRIME" OPERATION WAS PERFORMED WITH NO DIFFICULTIES NOTED. A REVIEW OF THE BLACK BOX AND PUMP HISTORIES SHOWS NO TIME/DATE RESETS OR DUPLICATE TIME/DATES IN THE RECORDED DATA. EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. INVESTIGATORS WERE UNABLE TO ADEQUATELY INVESTIGATE THE TIME LOSS/GAIN COMPLAINT DUE TO A DEFECTIVE BT1 FAILURE ON THE BENCH.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013, REPORTING THAT THE PUMP TIME RESETS ITSELF. THE PATIENT DENIED ANY TIME AND DATE RESET DURING BATTERY CHANGE. THE PATIENT STATED THAT THE TIME RANDOMLY RESETS ON ITS OWN. CUSTOMER SUPPORT INSTRUCTED THE PATIENT TO DISCONNECT AND REBOOT TO CHECK IF TIME AND DATE RESET AND IT DID NOT. THE PATIENT CONFIRMED THAT SHE NEVER HAD TO RESET THE TIME AND DATE HOWEVER, TWICE THE TIME WAS OFF BY 12 HOURS. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE TIME LOSS/GAIN ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58661 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |