FDA Adverse Event Injury Summary report: N

SECURA DR

MDR report key: 2955264 · Received February 11, 2013

Report

Report Number
3004209178-2013-01907
Event Type
Injury
Date Received
February 11, 2013
Report Date
February 15, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Removal / Correction Number
Z-0115-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS 5076 IMPLANTABLE PACING LEAD (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 99 % OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THE TIME OF RECOMMENDED REPLACEMENT TIME (RRT) WAS 2012 (B)(4). THE RRT VOLTAGE WAS LESS THAN OR EQUAL TO 2.625. THE WEEKLY BATTERY VOLTAGE TREND DATA SHOWED A MINIMUM BATTERY VOLTAGE OF 2.632 TO 2.618 VOLTS BETWEEN 2012 (B)(4) AND 2012 (B)(4). THERE WAS A LOW BATTERY ALERT ON 2012 (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)  HAD EARLY BATTERY DEPLETION. THE ICD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59316 SECURA DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224DRG

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 6949 IMPLANTABLE TACHY LEAD