FDA Adverse Event Malfunction Summary report: N

CONCERTO II CRT-D

MDR report key: 2955213 · Received February 11, 2013

Report

Report Number
3004209178-2013-01869
Event Type
Malfunction
Date Received
February 11, 2013
Report Date
November 26, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4194 IMPLANTABLE PACING LEAD 2010 (B)(6), 6947 IMPLANTABLE DEFIBRILLATION LEAD 2010 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION EPISODES THAT THE DEVICE HAD NOISE ON THE ATRIAL CHANNEL AND IT WAS OVERSENSED. THERE ALSO APPEARED TO BE SOME NOISE ON THE BASELINE FOR THE RIGHT VENTRICULAR (RV) LEAD CHANNEL BUT IT WAS NOT SENSED AND THE VENTRICULAR RHYTHM WAS PACED DURING THE EPISODES. THE DEVICE AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58242 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D274TRK

Patients

Seq Age Sex Outcome Treatment
1 1688TC COMPETITOR IMPLANTABLE PACING LEAD