ADJ WIRE COLLET 0.7-1.8MM
Report
- Report Number
- 0001811755-2013-00240
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 30, 2013
- Report Date
- January 30, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE QUALITY SPECIALIST CONFIRMED THE DEVICE HAD METAL SHAVINGS LOCATED ON THE DRIVESHAFT BY THE LEVER AND NOTED THE DRIVESHAFT HAD SCORE MARKS. ACCORDING TO A REVIEW OF THE RISK DOCUMENTS, WEAR ON THE DRIVESHAFT MAY PRODUCE METAL SHAVINGS. THEREFORE, IT IS LIKELY THE REPORTED EVENT WAS DUE TO DRIVESHAFT SCORING.
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE ADJ WIRE COLLET 0.7-1.8MM WAS SENT FOR SERVICE AND DURING FAILURE ANALYSIS IT WAS NOTED THAT THERE ARE METAL SHAVINGS PRESENT ON THE DRIVESHAFT OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
THE ADJ WIRE COLLET 0.7-1.8MM WAS SENT FOR SERVICE AND DURING FAILURE ANALYSIS IT WAS NOTED THAT THERE ARE METAL SHAVINGS PRESENT ON THE DRIVESHAFT OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59847 | ADJ WIRE COLLET 0.7-1.8MM | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | KIJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |