FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX MRI SURESCAN
MDR report key: 2954846
·
Received February 11, 2013
Report
- Report Number
- 2649622-2013-01351
- Event Type
- Injury
- Date Received
- February 11, 2013
- Date of Event
- May 17, 2012
- Report Date
- November 20, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, THE DISTAL ELECTRODE WAS COVERED IN BLOOD. THERE WAS APPARENT EXPLANT DAMAGE.
Additional Manufacturer Narrative · 1
(B)(4)
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD AND THE RIGHT ATRIAL (RA) LEADS DISLODGED. THE PATIENT HAD HAD AN ABLATION PROCEDURE A FEW MONTHS AFTER THE IMPLANT. THE PATIENT ALSO COMPLAINED OF THE DEVICE "MOVING AROUND" IN THE POCKET. BOTH LEADS WERE EXPLANTED AND REPLACED. THE DEVICE IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59915 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00069 YR | Hospitalization| R | (B)(4) IMPLANTABLE PULSE GENERATOR |