FDA Adverse Event Malfunction Summary report: N

XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 2954808 · Received February 11, 2013

Report

Report Number
2024168-2013-00720
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
November 24, 2012
Report Date
January 15, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. FAILURE TO ADVANCE/CROSS THE LESION COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. SHAFT SEPARATION WAS CONFIRMED. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. IT SHOULD BE NOTED THAT THE XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE STATES: SHOULD ANY RESISTANCE BE FELT AT ANY TIME DURING EITHER LESION ACCESS OR REMOVAL OF THE DELIVERY SYSTEM POST-STENT IMPLANTATION, THE ENTIRE SYSTEM SHOULD BE REMOVED AS A SINGLE UNIT. FAILURE TO FOLLOW THESE STEPS AND/OR APPLYING EXCESSIVE FORCE TO THE DELIVERY SYSTEM CAN POTENTIALLY RESULT IN LOSS OR DAMAGE TO THE STENT AND/OR DELIVERY SYSTEM COMPONENTS.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE MODERATELY CALCIFIED, MILDLY TORTUOUS, 99% STENOSED, PROXIMAL RIGHT CORONARY ARTERY. PREDILATATION OF THE LESION WAS PERFORMED PRIOR TO THE ATTEMPT TO STENT WITH A 2.75X18 MM XIENCE V STENT; HOWEVER, AFTER MULTIPLE ATTEMPTS WERE MADE TO CROSS THROUGH THE ANATOMY, THE PROXIMAL SHAFT OF THE STENT DELIVERY SYSTEM SEPARATED. ADDITIONAL PRE-DILATATION WAS PERFORMED AND A NEW XIENCE V STENT WAS DEPLOYED SUCCESSFULLY. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR ADVERSE PATIENT EFFECTS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59716 XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 1121241

Patients

Seq Age Sex Outcome Treatment
1 82 YR