FDA Adverse Event Malfunction Summary report: N

PICK 3722066 1MM PT 90 DEG

MDR report key: 2954802 · Received February 11, 2013

Report

Report Number
9680837-2013-00002
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 17, 2013
Report Date
January 17, 2013
Manufacturer
MEDTRONIC, INC.
Product Code
JYT
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. (B)(4). THE DEVICE WAS NOT RETURNED. THE DEVICE WAS NOT RETURNED AND THEREFORE NO EVALUATION COULD BE PERFORMED.

Additional Manufacturer Narrative · 1

THE COMPLAINT DEVICE WAS RETURNED FOR ANALYSIS. ANALYSIS FOUND THAT THE TIP WAS BROKEN OFF AT THE EXTREME END OF THE INSTRUMENT. AT THIS POINT IN THE SHAFT, THE DIAMETER OF METAL IS VERY THIN SUCH THAT A 90 DEGREE BEND IN THE METAL CAN BE MADE BY THE SUPPLIER AND LEAVE A 1.0 MM PICK EXTENSION. THE STRESS ON THE METAL TO PRODUCE A 90 DEGREE BEND IS SIGNIFICANT ON THE 1.4021 SURGICAL STAINLESS STEEL MATERIAL. THIS UNIT WAS MANUFACTURED IN JUNE 2012, SO AGE OF THE INSTRUMENT IS NOT A FACTOR. THERE WERE NO OTHER UNITS FROM THIS MANUFACTURING LOT TO REVIEW. COMPARING THIS UNIT TO THE SAME INSTRUMENT IN LOT 201206MF2 SHOWED NO DIFFERENCES. ANALYSIS HAS CONCLUDED THAT THE MOST LIKELY ROOT CAUSE IS DUE TO EXCESSIVE FORCE BEING APPLIED TO THE END OF THE DEVICE.

Additional Manufacturer Narrative · 1

IN A PREVIOUS MDR, AN EXPIRATION DATE WAS REPORTED. THIS DEVICE DOES NOT HAVE AN EXPIRATION DATE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, THE PICK'S TIP BROKE OFF IN THE CHILD'S EAR. THE DOCTOR WAS PERFORMING A COCHLEAR IMPLANT ON THE CHILD AND SAW THAT THE TIP BROKE OFF. HE STATED THAT HE SAW THE SUCTION SUCK UP THE TIP OF THE PICK. THE DOCTOR IS FULLY CONFIDENT THAT THE TIP WAS SUCKED UP BY THE SUCTION, AND THE DOCTOR IS NOT GOING TO DO AN X-RAY. THERE WAS NO PATIENT IMPACT AND THERE WAS NO DELAY IN SURGERY. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59714 PICK 3722066 1MM PT 90 DEG PICK, MICROSURGICAL EAR JYT MEDTRONIC, INC. 3722066 201206MF1

Patients

Seq Age Sex Outcome Treatment
1 00002 YR