FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2954801 · Received February 11, 2013

Report

Report Number
2649622-2013-02401
Event Type
Injury
Date Received
February 11, 2013
Date of Event
November 1, 2012
Report Date
May 30, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT: 5076 IMPLANTABLE PACING LEAD (B)(6) 2008. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #(B)(4). THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THERE WERE NO ANOMALIES OBSERVED ON THE RETURNED LEAD PROXIMAL SEGMENT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. THE FULL LEAD WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) EXHIBITED HIGH IMPEDANCE MEASUREMENTS. THE LEAD REMAINS IN SERVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) EXHIBITED HIGH IMPEDANCE MEASUREMENTS. THE LEAD REMAINS IN SERVICE. IT WAS FURTHER REPORTED THAT THE RV LEAD HAD AN APPARENT FRACTURE AND WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58675 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR