FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2954769
·
Received February 11, 2013
Report
- Report Number
- 2649622-2013-02393
- Event Type
- Injury
- Date Received
- February 11, 2013
- Date of Event
- November 29, 2012
- Report Date
- November 29, 2012
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. (B)(6). (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE LEAD IMPLANT HIGH THRESHOLD AND HIGH IMPEDANCE WERE NOTED. THE LEAD WAS LATER EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58970 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00082 YR | Hospitalization| R |