FDA Adverse Event Malfunction Summary report: N

CONTOUR CURVED CUTTER STAPLER

MDR report key: 2954716 · Received February 11, 2013

Report

Report Number
3005075853-2013-00572
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 23, 2013
Report Date
January 25, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE DEVICE DID NOT STAPLE ON THE FIRST FIRING. THE SURGEON SUTURED TO COMPLETE THE PROCEDURE. PROCEDURE WAS PROLONGED 25 MINUTES. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59064 CONTOUR CURVED CUTTER STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1