FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 2954697 · Received February 11, 2013

Report

Report Number
3004209178-2013-02479
Event Type
Injury
Date Received
February 11, 2013
Report Date
November 14, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT PRODUCTS: 4076 IMPLANTABLE PACING LEAD: (B)(6) 2007. 6949 IMPLANTABLE TACHY LEAD: (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE MAY HAVE REACHED THE ELECTIVE REPLACEMENT INDICATOR (ERI) PREMATURELY DUE TO HIGH LEFT VENTRICULAR ( LV) LEAD THRESHOLDS. IT WAS ALSO REPORTED THAT THE ATRIAL LEAD HAD HIGH THRESHOLD MEASUREMENTS. THE DEVICE WAS EXPLANTED AND REPLACED; THE LV AND ATRIAL LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58658 MAXIMO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284TRK

Patients

Seq Age Sex Outcome Treatment
1 00065 YR Hospitalization| R 5071 IMPLANTABLE PACING LEAD