FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 2954484 · Received February 11, 2013

Report

Report Number
3004209178-2013-02407
Event Type
Injury
Date Received
February 11, 2013
Report Date
March 15, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031/S084
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED NORMAL BATTERY DEPLETION. (B)(4). (B)(6).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEADS, (B)(6) 2009; IMPLANTABLE PACING LEAD, (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD EARLY BATTERY DEPLETION. IT WAS NOTED THAT THE LEFT VENTRICULAR LEAD HAD HIGH THRESHOLDS WHICH DECREASED LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEENREPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58552 MAXIMO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284TRK

Patients

Seq Age Sex Outcome Treatment
1 00093 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD