FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2954468 · Received February 11, 2013

Report

Report Number
2649622-2013-02241
Event Type
Injury
Date Received
February 11, 2013
Date of Event
December 13, 2012
Report Date
December 27, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S024
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED, HELIX EXTENSION/RETRACTION AND LENGTH TESTING WERE PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOLLOWING THE IMPLANT PROCEDURE AN X-RAY SHOWED THE RIGHT VENTRICULAR LEAD WAS DISLODGED. AN ATTEMPT TO REPOSITION THE LEAD WAS UNSUCCESSFUL. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58277 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00055 YR Hospitalization| R