FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 2954438 · Received February 11, 2013

Report

Report Number
9614453-2013-00454
Event Type
Injury
Date Received
February 11, 2013
Date of Event
November 2, 2012
Report Date
November 2, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). EVALUATION SUMMARY: (B)(4): REVIEW OF PERFORMANCE DATA SHOWED THAT THE DEVICE WAS APPROACHING RRT (RECOMMENDED REPLACEMENT TIME). UPON FULL ANALYSIS, NORMAL BATTERY DEPLETION WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DID NOT REACH EXPECTED LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58185 MAXIMO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D284TRK

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Hospitalization| R