FDA Adverse Event
Injury
Summary report: N
SIGMA 300 SR
MDR report key: 2954420
·
Received February 11, 2013
Report
- Report Number
- 9614453-2013-00456
- Event Type
- Injury
- Date Received
- February 11, 2013
- Report Date
- December 27, 2012
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE REVEALED LIFTED HYBRID BOND WIRES.
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58054 | SIGMA 300 SR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | SSR303 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |