FDA Adverse Event Injury Summary report: N

SIGMA 300 SR

MDR report key: 2954420 · Received February 11, 2013

Report

Report Number
9614453-2013-00456
Event Type
Injury
Date Received
February 11, 2013
Report Date
December 27, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035/S002
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE REVEALED LIFTED HYBRID BOND WIRES.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THEY WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58054 SIGMA 300 SR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND SSR303

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R