FDA Adverse Event Malfunction Summary report: N

DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE

MDR report key: 2954418 · Received February 11, 2013

Report

Report Number
2649622-2013-02313
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
December 21, 2012
Report Date
December 20, 2012
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THERE WERE NO ANOMALIES FOUND. IT WAS NOTED THAT THERE WAS BLOOD COVERING THE DISTAL END. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE RIGHT VENTRICULAR (RV) LEAD COULD NOT BE ROTATED UNDER THE CONDITION THAT THE HELIX WAS NOT EXTENDED. IT WAS NOTED THAT THE LEAD ALSO HAD HIGH IMPEDANCE, EVEN AFTER REPOSITION THREE TIMES. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED INSTEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58075 DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00070 YR