ATTAIN OTW
Report
- Report Number
- 2649622-2013-02229
- Event Type
- Injury
- Date Received
- February 11, 2013
- Date of Event
- December 13, 2012
- Report Date
- December 13, 2012
- Manufacturer
- MPRI
- Product Code
- LWP
- PMA / PMN Number
- P010015/S012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT AND DEVELOPED COSMETIC ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IT WAS REPORTED THAT DURING THE PROPHYLACTIC EXTRACTION OF THE RIGHT VENTRICULAR (RV) LEAD, THE LEFT VENTRICULAR (LV) LEAD WAS INADVERTENTLY DISLODGED. ANOTHER LV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59727 | ATTAIN OTW | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MPRI | 419488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00080 YR | 6949 IMPLANTABLE TACHY LEAD |