FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2954345 · Received February 11, 2013

Report

Report Number
2649622-2013-02285
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
November 8, 2012
Report Date
November 8, 2012
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX WAS OUT OF SPECIFICATION AS IT TOOK TOO MANY TURNS TO EXTEND. IT WAS ALSO NOTED THAT THE DISTAL ELECTRODE WAS DAMAGED AND THE DISTAL CONDUCTOR WAS COVERED IN BLOOD (NOT OBSTRUCTED). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT THE LEAD HAD POOR THRESHOLD AND POOR SENSING VALUES. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58766 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00067 YR