FDA Adverse Event
Injury
Summary report: N
MAXIMO II CRT-D
MDR report key: 2954332
·
Received February 11, 2013
Report
- Report Number
- 3004209178-2013-02395
- Event Type
- Injury
- Date Received
- February 11, 2013
- Report Date
- March 22, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS CONTINUED: 694765 IMPLANTABLE TACHY LEAD: (B)(6) 2009. 419488 IMPLANTABLE PACING LEAD: (B)(6) 2009. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS A POTENTIAL PERFORMANCE ISSUE WITH THE DEVICE AS ITS LONGEVITY WAS LESS THAN FOUR YEARS WHEN IT REACHED ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58324 | MAXIMO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D284TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Hospitalization| R | 507652 IMPLANTABLE PACING LEAD |