FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2954330 · Received February 11, 2013

Report

Report Number
3004209178-2013-02406
Event Type
Injury
Date Received
February 11, 2013
Report Date
November 26, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031/S084
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE DID NOT DETECT ANY PERFORMANCEISSUES, BUT THE CALCULATED LONGEVITY RESULT OF 82% OF EXPECTED WAS LESS THAN THE PREDICTED VALUE BASED ON THE AVAILABLE PROGRAMMED PARAMETERS. IN THE ABSENCE OF THE COMPLETE PROGRAMMING HISTORY, IT IS NOT POSSIBLE TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4968 X 2 IMPLANTABLE PACING LEADS (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD UNEXPECTED LONGEVITY AND REACHED RECOMMENDED REPLACEMENT TIME (RRT) IN UNDER 5 YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58683 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00087 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD