ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-00546
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 28, 2013
- Report Date
- January 29, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE A WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED - EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE 'INSTRUMENT ERROR' ALERT WAS DISPLAYED; AND WHEN TESTED ON THE GEN04 GENERATOR, AN ERROR CODE 5 (INSTRUMENT ERROR) WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE TO STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 OR INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS 'REPLACE INSTRUMENT' WITH THE GEN11 LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE. DEVICE B WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED WITH A GENERATOR AND WAS FUNCTIONAL. IN ADDITION, THE DEVICE ACTIVATED WITH BOTH MAX AND MIN BUTTONS. THE TISSUE PAD OF THE DEVICE WAS IN GOOD PHYSICAL CONDITION. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. IT COULD NOT BE DETERMINED WHY A NOISE REPORTEDLY OCCURRED FROM THE SHAFT AND AN ERROR WAS DISPLAYED. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY, A NOISE OCCURRED FROM THE SHAFT AND AN ERROR WAS DISPLAYED. AN ERROR CODE WAS UNKNOWN. THE BLADE WAS BROKEN OFF WHEN THE DEVICE WAS ACTIVATED AGAIN. THE BROKEN BLADE FELL OUTSIDE PATIENT'S BODY. NO PIECES FELL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. TWO DEVICES WILL BE RETURNING. ONE OF THEM WAS .
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58507 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | J92UOH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |