RETROFLEX3 DELIVERY SYSTEM
Report
- Report Number
- 2015691-2013-19244
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- December 6, 2012
- Report Date
- February 6, 2013
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPT
- PMA / PMN Number
- P110021
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE RETROFLEX DELIVERY SYSTEM WAS RETURNED TO EDWARDS LIFESCIENCES FOR EVALUATION. A MORE SIGNIFICANT FAILURE THAN WAS ORIGINALLY REPORTED WAS DETECTED AT THIS TIME. VISUAL AND FUNCTIONAL TESTING WAS PERFORMED. UNDER MAGNIFICATION, ONE OF THE ADHESIVE PORTS APPEARED CONCAVE. DURING FUNCTIONAL TESTING, THE DEVICE FAILED THE PRESSURE DECAY LEAK TEST. IN ADDITION, THE DELIVERY BALLOON COULD NOT BE SIZED WITH THE SIZING RING AS THE DELIVERY SYSTEM WAS LEAKING THROUGH THE GUIDEWIRE FLUSH PORT. THIS WAS FOLLOWED BY A RED DYE TEST. AS THE BALLOON WAS INFLATED THROUGH THE INFLATION PORT, A LEAK WAS VISIBLE THROUGH THE GUIDE-WIRE LUMEN. POST TESTING, VISUAL INSPECTION OF THE Y-CONNECTOR REVEALED INDICATION OF A LEAK PATH AT THE Y-CONNECTOR, BETWEEN THE BALLOON PORT AND GUIDEWIRE PORT, THROUGH THE ADHESIVE, AROUND THE GUIDEWIRE LUMEN. THE COMPLAINT WAS CONFIRMED AS THE DELIVERY SYSTEM WAS UNABLE TO BE DE-AIRED AND A LEAK PATH WAS OBSERVED AT THE Y-CONNECTOR. THE ROOT CAUSE OF THE LEAK IN THE DELIVERY SYSTEM WAS DETERMINED TO BE A LEAK PATH IN THE Y-CONNECTOR TO THE GUIDE-WIRE SHAFT BOND. DURING EVALUATION ONE COULD SEE THE LEAK PATH WHERE THE SALINE AND CONTRAST SOLUTION ENTERED DUE TO THE ABSENCE OF ADHESIVE IN THAT AREA. PER THE MANUFACTURING PROCEDURE, A 100% LEAK TEST IS PERFORMED. IN ADDITION, THE Y-CONNECTOR BONDS ARE 100% INSPECTED FOR VOIDS, BUBBLES, OR VISUAL DEFECTS AND THE PROCEDURE STATE TO REJECT ANY COMPONENTS WHERE THE ADHESIVE IS COVERING LESS THAN 90% OF THE BONDS SURFACE, OR IF THERE IS ANY VISUAL INDICATION OF LEAK PATH, EXCESSIVE ADHESIVE ON THE OUTSIDE OF THE Y-CONNECTOR, EVIDENCE OF ADHESIVE WITHIN THE OUTER SHAFT OR GW LUMENS, PARTICULATE CAPTURE, BURNS, OR OTHER VISUAL DEFECTS. THESE INSPECTIONS MAKE IT UNLIKELY THAT A PRE-EXISTING LEAK PATH WAS PRESENT DURING THE MANUFACTURING PROCESS. IN THIS CASE, IT IS DIFFICULT TO DETERMINE IF THE LEAK PATH CONDITION OF THIS SPECIFIC UNIT WAS MISSED IN MANUFACTURING OR IF IT DEVELOPED AT SOME POINT AFTER COMPLETION OF THE MANUFACTURING PROCESS. RETRAINING OF THE MANUFACTURING OPERATORS WAS PERFORMED TO REJECT FOR LEAK PATHS IN THE Y-CONNECTOR TO GUIDEWIRE LUMEN BOND AND TO HAVE A MINIMUM OF 90% ADHESIVE COVERAGE IN THE BOND AREA. THE INSTRUCTIONS FOR USE AND THE TRAINING MANUAL WERE REVIEWED AND NO IFU AND TRAINING DEFICIENCIES WERE IDENTIFIED.THE IFU STEPS FOR PREPPING THE RF3 DELIVERY SYSTEM CALL FOR THE DEVICE FIRST TO BE DE-AIRED AND THEN TO SIZE THE BALLOON BY INFLATING THE BALLOON INSIDE THE BALLOON GAUGE UNTIL THE BALLOON REACHES THE CORRECT DIAMETER. AFTER THE BALLOON IS PROPERLY SIZED, THE BALLOON IS DEFLATED. THE PREPPING AND BALLOON SIZING STEPS MAKE IT HIGHLY LIKELY THAT A Y-CONNECTOR LEAK PATH WOULD BE DETECTED BEFORE EVER REACHING A PATIENT. NO FURTHER ACTION IS REQUIRED AT THIS TIME. TRENDING FOR THIS ISSUE WILL CONTINUE TO BE MONITORED.
AS REPORTED BY THE EDWARDS CLINICAL SPECIALIST, THE RETROFLEX3 DELIVERY SYSTEM LEAKED THROUGH THE WIRE PORT DURING PREPARATION OF THE DEVICE PRIOR TO USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 58393 | RETROFLEX3 DELIVERY SYSTEM | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | EDWARDS LIFESCIENCES | 9120FS23 | 59250507 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |