FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2953945 · Received January 15, 2013

Report

Report Number
3008642652-2013-00109
Event Type
Malfunction
Date Received
January 15, 2013
Date of Event
December 12, 2012
Report Date
January 9, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK TE PAD MESSAGES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THERE WAS AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG ELECTRODES A AND B. THE CAUSE OF THE CHECK THE PAD MESSAGES IS THE OPEN PULSE WIRE. THE ROOT CAUSE OF THE OPEN PULSE WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK THE PAD MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21892 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 39 YR