CONSULTA CRT-D
Report
- Report Number
- 3004209178-2013-02042
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- February 19, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 6947 IMPLANTABLE TACHY LEAD: (B)(6) 2009. 4196 IMPLANTABLE PACING LEAD: (B)(6) 2009. (B)(4).
PRODUCT EVENT SUMMARY # THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. PERFORMANCE DATA COLLECTED FROM THE DEVICE REVEALED THAT THE DEVICE WENT TO ELECTIVE REPLACEMENT INDICATOR (ERI) ON (B)(6) 2012.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD UNEXPECTED LONGEVITY AND REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) WITHIN THE WARRANTY PERIOD. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLDS. THE DEVICE WAS EXPLANTED AND REPLACED AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56690 | CONSULTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D224TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00065 YR | Hospitalization| R | 3830 IMPLANTABLE PACING LEAD |