MAXIMO II CRT-D
Report
- Report Number
- 9614453-2013-00264
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- November 14, 2012
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.(B)(4). PRODUCT EVENT SUMMARY: PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. RECOMMENDED REPLACEMENT TIME (RRT) IN SAVE-TO-DISK OCCURRED ON (B)(4) 2012; DEVICE RRT<(><<)>=2.6251 VOLTS. THE WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.629 TO 2.602 VOLTS BETWEEN (B)(4) 2012. THERE WERE TWO LOW BATTERY VOLTAGE ALERTS (B)(4) 2012.
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
IT WAS REPORTED THAT THE BATTERY DEPLETED WITHIN THREE YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56504 | MAXIMO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | IPG MFG SWITZERLAND | D284TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |