FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 2953732 · Received February 9, 2013

Report

Report Number
9614453-2013-00264
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 14, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.(B)(4). PRODUCT EVENT SUMMARY: PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. RECOMMENDED REPLACEMENT TIME (RRT) IN SAVE-TO-DISK OCCURRED ON (B)(4) 2012; DEVICE RRT<(><<)>=2.6251 VOLTS. THE WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.629 TO 2.602 VOLTS BETWEEN (B)(4) 2012. THERE WERE TWO LOW BATTERY VOLTAGE ALERTS (B)(4) 2012.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY DEPLETED WITHIN THREE YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56504 MAXIMO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D284TRK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R