FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2953714 · Received February 9, 2013

Report

Report Number
2649622-2013-01328
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 21, 2012
Report Date
November 21, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE HELIX DEPLOYED NORMALLY. DURING THE LEAD PLACEMENT, THERE WERE SEVERAL DISLODGEMENTS RESULTING IN SEVERAL REPOSITIONINGS. THEN THE HELIX WOULD NOT RETRACT AFTER TWENTY-FIVE TO THIRTY TURNS WHILE IN THE BODY. THE PHYSICIAN REMOVED THE LEAD AND ONCE OUTSIDE THE BODY THE HELIX RETRACTED. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56392 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00092 YR