FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 2953669 · Received February 9, 2013

Report

Report Number
3004209178-2013-02015
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 29, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED MULTIPLE SHOCKS FOR T-WAVE OVERSENSING (TWOS) DURING EXERCISE. THE DEVICE WAS EXPLANTED AND REPLACED WITH A DEVICE THAT HAS A T-WAVE ALGORITHM, AND THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57123 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO OPERATIONS CO., JUNCOS D154VWC

Patients

Seq Age Sex Outcome Treatment
1 00049 YR Hospitalization| L| R 6947 IMPLANTABLE TACHY LEAD