FDA Adverse Event Injury Summary report: N

ELECTRODE, PACEMAKER, PERMANENT

MDR report key: 2953601 · Received February 9, 2013

Report

Report Number
2649622-2013-01178
Event Type
Injury
Date Received
February 9, 2013
Report Date
February 28, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P830061
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD A PERFORMANCE ISSUE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONSHAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD ISSUE WAS THAT THE THRESHOLDS SUDDENLY INCREASED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56170 ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4092-58

Patients

Seq Age Sex Outcome Treatment
1 00091 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR (IPG)