FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2953548 · Received February 9, 2013

Report

Report Number
2649622-2013-01212
Event Type
Injury
Date Received
February 9, 2013
Report Date
November 20, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY #(B)(4) - THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED COSMETIC OUTER INSULATION DEPRESSION. THE OUTER INSULATION WAS BREACH CUT. THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EXPLANT OF THE DEVICE, THE RIGHT VENTRICULAR (RV) LEAD WAS DAMAGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EXPLANT OF THE DEVICE, THE RIGHT VENTRICULAR (RV) LEAD WAS DAMAGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55932 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Hospitalization| R