FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2953504 · Received February 9, 2013

Report

Report Number
2649622-2013-01134
Event Type
Injury
Date Received
February 9, 2013
Date of Event
October 31, 2012
Report Date
November 1, 2012
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO T-WAVE OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56640 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Life Threatening (B)(4) IMPLANTABLE CARDIAC DEFIBRILLATOR