FDA Adverse Event
Injury
Summary report: N
TBD
MDR report key: 2953444
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-01113
- Event Type
- Injury
- Date Received
- February 9, 2013
- Date of Event
- December 3, 2012
- Report Date
- December 3, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS APPARENT EXPLANT DAMAGE.
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT IMPLANT THE ATRIAL TINED LEAD HAD MARGINAL THRESHOLDS AFTER A DIFFICULT CASE. TWO DAYS LATER THE THRESHOLDS HAD NOT IMPROVED DUE TO POOR LEAD POSITION SO THE LEAD WAS EXPLANTED AND AN ACTIVE FIXATION LEAD WAS PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56040 | TBD | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5592-53 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00087 YR | Hospitalization| R | (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |