FDA Adverse Event Injury Summary report: N

TBD

MDR report key: 2953444 · Received February 9, 2013

Report

Report Number
2649622-2013-01113
Event Type
Injury
Date Received
February 9, 2013
Date of Event
December 3, 2012
Report Date
December 3, 2012
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT IMPLANT THE ATRIAL TINED LEAD HAD MARGINAL THRESHOLDS AFTER A DIFFICULT CASE. TWO DAYS LATER THE THRESHOLDS HAD NOT IMPROVED DUE TO POOR LEAD POSITION SO THE LEAD WAS EXPLANTED AND AN ACTIVE FIXATION LEAD WAS PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56040 TBD ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5592-53

Patients

Seq Age Sex Outcome Treatment
1 00087 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR