CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2013-01090
- Event Type
- Injury
- Date Received
- February 9, 2013
- Date of Event
- December 3, 2012
- Report Date
- December 3, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN THE DISTAL ELECTRODE, AND THERE WAS APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076, IMPLANTABLE PACING LEAD, (B)(6) 2012. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING AN UPGRADE PROCEDURE THE PHYSICIAN ATTEMPTED TO MOVE THE RIGHT ATRIAL (RA) LEAD TO GET BETTER THRESHOLDS, AND AFTER MOVING THE LEAD SEVERAL TIMES THE HELIX WOULD NO LONGER EXTEND FULLY. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57513 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00087 YR | Hospitalization| R | VEDR01 IMPLANTABLE PULSE GENERATOR (IPG) |