PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2182208-2013-00312
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Report Date
- November 14, 2012
- Manufacturer
- RICE CREEK MFG
- Product Code
- LWP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; CASE HALVES AND WHEEL ASSEMBLIES ARE BROKEN OFF AND MISSING. CABLE IS TWISTED APPROXIMATELY 19 INCHES FROM HEART WIRE CONNECTOR AND APPROXIMATELY 28 INCHES FROM EPG (EXTERNAL PULSE GENERATOR) PLUG. CABLE CONTINUITY TESTS OK. NO INTERMITTENT CONNECTION FOUND WHEN CABLE IS BENT /MANIPULATED. CONNECTIONS WERE NOT SHORTING OUT BETWEEN THEMSELVES.
(B)(4). THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. WITHOUT A DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED.
THE CABLES WERE RETURNED.
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) CABLES WERE "BROKEN." OF NOTE, THE PLASTIC SCREW KNOB SHOWED "PHYSICAL BREAKAGE." THE HOSPITAL HAD RECENTLY REVISED THEIR CLEANING/STERILIZATION PROCESS AND NOTED THE CABLES WERE "BREAKING" FOLLOWING THE STERILIZATION PROCESS. THE CABLES WILL BE RETURNED. NO PATIENT INVOLVEMENT OR COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56886 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | RICE CREEK MFG | 5433AL | 0815 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |