FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
MDR report key: 2953314
·
Received February 9, 2013
Report
- Report Number
- 2182208-2013-00319
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- December 29, 2012
- Report Date
- December 29, 2012
- Manufacturer
- RICE CREEK MANUFACTURING
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE KEYBOARD WAS MISSING, CAUSING THE LONG BOOT SEQUENCE. WHEN THE KEYBOARD WAS REPLACED THE BOOT SEQUENCE WAS NORMAL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PROGRAMMER'S BOOT-UP TIME EXCEEDED 90 SECONDS AND THAT IT KEPT REBOOTING. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56419 | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MANUFACTURING | 2090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |