FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 2953314 · Received February 9, 2013

Report

Report Number
2182208-2013-00319
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
December 29, 2012
Report Date
December 29, 2012
Manufacturer
RICE CREEK MANUFACTURING
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE KEYBOARD WAS MISSING, CAUSING THE LONG BOOT SEQUENCE. WHEN THE KEYBOARD WAS REPLACED THE BOOT SEQUENCE WAS NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER'S BOOT-UP TIME EXCEEDED 90 SECONDS AND THAT IT KEPT REBOOTING. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56419 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MANUFACTURING 2090

Patients

Seq Age Sex Outcome Treatment
1