MAXIMO II CRT-D
Report
- Report Number
- 3004209178-2013-01912
- Event Type
- Injury
- Date Received
- February 9, 2013
- Report Date
- February 7, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY : THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT : 6947, IMPLANTABLE TACHY LEAD, (B)(6) 2011; 5076, IMPLANTABLE PACING LEAD, (B)(6) 2011. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DEVICE HAD POSSIBLE PREMATURE BATTERY DEPLETION. IT WAS ALSO NOTED THAT THE VENTRICULAR PACING OUTPUTS ONTHE DEVICE WERE PROGRAMMED HIGH DUE TO CHRONIC HIGH THRESHOLD ON THE COMPETITOR LEFT VENTRICULAR LEAD. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55775 | MAXIMO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D284TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00072 YR | Hospitalization| R | 4047 COMPETITOR IMPLANTABLE PACING LEAD |