VIRTUOSO DR
Report
- Report Number
- 9614453-2013-00212
- Event Type
- Injury
- Date Received
- February 9, 2013
- Date of Event
- October 31, 2012
- Report Date
- November 21, 2012
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. CONCOMITANT PRODUCTS: 0154 COMPETITOR IMPLANTABLE LEAD: (B)(6) 2002. (B)(4).
IT WAS REPORTED THAT THE PATIENT HAD A PATIENT ALERT FOR RECOMMENDED REPLACEMENT TIME (RRT) OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). PREMATURE DEPLETION WAS SUSPECTED. THE ICD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55670 | VIRTUOSO DR | LWS | IPG MFG SWITZERLAND | D164AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00053 YR | Hospitalization| R | 48-35S COMPETITOR IMPLANTABLE LEAD |