FDA Adverse Event Injury Summary report: N

CONCERTO II CRT-D

MDR report key: 2953211 · Received February 9, 2013

Report

Report Number
3004209178-2013-01945
Event Type
Injury
Date Received
February 9, 2013
Report Date
December 3, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE LOST TELEMETRY AND FUNCTION DUE TO BATTERY DEPLETION. THE CAUSE OF THE DEPLETION CANNOT BE DETERMINED. THE DEVICE WAS FULLY FUNCTIONAL AND OPERATED UNDER NORMAL CURRENT DRAIN WHEN POWERED WITH AN EXTERNAL SUPPLY. THE RESIDUAL VOLTAGE AND IMPEDANCE INDICATES THAT THE BATTERY WAS NOT INTERNALLY SHORTED.. SINCE NO SAVE TO DISK DATA COULD BE RETRIEVED FROM THE DEVICE AND NO OTHER DATA WAS PROVIDED FROM THE FIELD THERE IS NO WAY TO CONFIRM THIS RESULT. THE RESULT OF ANALYSIS IS INCONCLUSIVE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 5076 IMPLANTABLE PACING LEAD, (B)(6) 2006; 4194 IMPLANTABLE PACING LEAD, (B)(6) 2006. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXHIBITED A SHORT BATTERY LIFE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXHIBITED A SHORT BATTERY LIFE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55679 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D274TRK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 6948 IMPLANTABLE TACHY LEAD