FDA Adverse Event Injury Summary report: N

BURKHART ROENTGEN RADIATION SHIELD

MDR report key: 295318 · Received September 1, 2000

Report

Report Number
MW4002840
Event Type
Injury
Date Received
September 1, 2000
Date of Event
August 14, 2000
Report Date
September 1, 2000
Manufacturer
BURKHART ROENTGEN INT'L, INC.
Product Code
IWQ
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

REPORTEDLY, DURING A PROCEDURE, THE RADIATION SHIELD DETACHED FROM ITS SUSPENSION ARM AND FELL, STRIKING THE PT'S ARM AND THE RADIOLOGIST'S WRIST. THE RADIOLOGIST SUSTAINED A HAIRLINE FRACTURE OF HIS WRIST. THE EXTEND OF INJURY, IF ANY, TO THE PT IS UNK AT THIS TIME. REPORTEDLY, THE RADIATIONSHIELD FELL DUE TO A BROKEN WELD ON THE YOKE WHICH HOLDS THE SHIELD ON THE SUSPENSION ARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BURKHART ROENTGEN RADIATION SHIELD PERSONNEL PROTECTIVE SHIELD IWQ BURKHART ROENTGEN INT'L, INC. BK2001 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention