FDA Adverse Event
Injury
Summary report: N
BURKHART ROENTGEN RADIATION SHIELD
MDR report key: 295318
·
Received September 1, 2000
Report
- Report Number
- MW4002840
- Event Type
- Injury
- Date Received
- September 1, 2000
- Date of Event
- August 14, 2000
- Report Date
- September 1, 2000
- Manufacturer
- BURKHART ROENTGEN INT'L, INC.
- Product Code
- IWQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
REPORTEDLY, DURING A PROCEDURE, THE RADIATION SHIELD DETACHED FROM ITS SUSPENSION ARM AND FELL, STRIKING THE PT'S ARM AND THE RADIOLOGIST'S WRIST. THE RADIOLOGIST SUSTAINED A HAIRLINE FRACTURE OF HIS WRIST. THE EXTEND OF INJURY, IF ANY, TO THE PT IS UNK AT THIS TIME. REPORTEDLY, THE RADIATIONSHIELD FELL DUE TO A BROKEN WELD ON THE YOKE WHICH HOLDS THE SHIELD ON THE SUSPENSION ARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BURKHART ROENTGEN RADIATION SHIELD | PERSONNEL PROTECTIVE SHIELD | IWQ | BURKHART ROENTGEN INT'L, INC. | BK2001 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Required Intervention |