CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2013-00885
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 30, 2012
- Report Date
- November 30, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. HELIX EXTENSION/RETRACTION AND LENGTH TESTING WERE PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS. PINCH ON TOOL THAT WAS RETURNED WITH LEAD WAS USED FOR TESTING.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT PRIOR TO IMPLANT OF THE LEAD THE HELIX WAS NOT ABLE TO EXTRACTED. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57037 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |