FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2953055 · Received February 9, 2013

Report

Report Number
2649622-2013-00870
Event Type
Injury
Date Received
February 9, 2013
Date of Event
November 21, 2012
Report Date
March 5, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND THAT THE DISTAL CONDUCTOR WAS FRACTURED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD TRIGGERED A LEAD INTEGRITY ALERT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57857 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR