FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO

MDR report key: 2952925 · Received February 9, 2013

Report

Report Number
2649622-2013-00762
Event Type
Malfunction
Date Received
February 9, 2013
Date of Event
November 29, 2012
Report Date
November 29, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX WAS OVER-RETRACTED, AND THE LEAD WAS DAMAGED AT IMPLANT.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD THE HELIX WOULD NOT EXTEND. HELIX EXTENSION COULD NOT BE SEEN ON FLUOROSCOPY AFTER 30 TURNS. UPON REMOVAL FROM THE BODY THE HELIX STILL WOULD NOT EXTEND AFTER 40 TURNS. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57320 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00068 YR