FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2952844
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-00736
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 29, 2012
- Report Date
- November 29, 2012
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED).
Additional Manufacturer Narrative · 1
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE LEAD'S HELIX WOULD NOT SCREW OUT. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 55916 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00084 YR |