FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2952841 · Received February 9, 2013

Report

Report Number
2182208-2013-00249
Event Type
Malfunction
Date Received
February 9, 2013
Report Date
November 20, 2012
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE DEVICE FAILED TELEMETRY TESTING DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE STYLUS PEN ON THE PROGRAMMER WAS NOT WORKING. THE PEN WAS REPLACED BUT IT WAS STILL NOT WORKING. THE PROGRAMMER AND RADIOFREQUENCY (RF) HEAD WERE RETURNED FOR SERVICE. BOTH PRODUCTS WERE ANALYZED AND TESTED OUT OF SPECIFICATION. THERE WAS NO INDICATION OF ANY PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55920 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2067

Patients

Seq Age Sex Outcome Treatment
1