FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 2952646
·
Received February 9, 2013
Report
- Report Number
- 2649622-2013-00648
- Event Type
- Malfunction
- Date Received
- February 9, 2013
- Date of Event
- November 6, 2012
- Report Date
- November 19, 2012
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 1882TC COMPETITOR IMPLANTABLE PACING LEAD (B)(6) 2012. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED WITHIN TWO WEEKS POST IMPLANT THAT THE PATIENT PRESENTED TO THE CLINIC WITH "HICCUP" SYMPTOMS. INTERROGATION SHOWED THERE WAS DOUBLE COUNTING WITH POTENTIAL FAR FIELD P-WAVE OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57812 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M62 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00061 YR | D314DRM IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |